Reduction of polarimetric data using Mueller calculus applied to Nasmyth instruments
Abstract
We present a method based on Mueller calculus to calibrate linear polarimetric observations. The key advantages of the proposed way of calibration are: (1) that it can be implemented in a data reduction pipeline, (2) that it is possible to do accurate polarimetry also for telescopes/instruments with polarimetric non-friendly architecture (e.g. Nasmyth instruments) and (3) that the proposed strategy is much less time consuming than standard calibration procedures. The telescope/instrument will polarimetrically be described by a train of Mueller matrices. The components of these matrices are dependent on wavelength, incident angle of the incoming light and surface properties.
Explore related subjects
Keep this discovery
Franco Joos, Esther Buenzli, Hans Martin Schmid, Christian Thalmann. 2008-12-08. Reduction of polarimetric data using Mueller calculus applied to Nasmyth instruments. https://doi.org/10.1117/12.788915