arXiv · 1307.0931
A novel scheme for measuring the relative phase difference between S and P polarization in optically denser medium
Abstract
We demonstrate applications of a novel setup which is used for measuring relative phase difference between S and P polarization at oblique incidence point in optically denser medium by analyzing the relative frequency shift of adjacent axial modes of S and P resonances of a monolithic folded Fabry-Perot cavity (MFC). The relative frequency shift of adjacent axial modes of S and P resonances of MFC is around 1.030 GHz for the confocal MFC cavity and 0.3869 GHz for the parallel MFC cavity respectively. The relative phase difference at reflection point A in optically denser medium is inferred to be around -167.4° for confocal cavity and -201.1° for parallel cavity separately. This scheme can be used to measure the relative phase difference in optically denser mediums which is out of measuring range of Ellipsometry.
Explore related subjects
Keep this discovery
Yu Peng. 2013-07-03. A novel scheme for measuring the relative phase difference between S and P polarization in optically denser medium. https://doi.org/10.1088/1612-2011%2F10%2F7%2F075006