arXiv · 1312.1297
System analysis of Force Feedback Microscopy
Abstract
It was shown recently that the Force Feedback Microscope can avoid the jump-to-contact in Atomic force Microscopy even when the cantilevers used are very soft, thus increasing force resolution. In this letter, we explore theoretical aspects of the associated real time control of the tip position. We take into account lever parameters such as the lever characteristics in its environment, spring constant, mass, dissipation coefficient and the operating conditions such as controller gains, and interaction force. We show how the controller parameters are determined so that the FFM functions at its best and estimate the bandwidth of the system under these conditions.
Explore related subjects
Keep this discovery
Mario S Rodrigues, Luca Costa, Joel Chevrier, Fabio Comin. 2013-11-20. System analysis of Force Feedback Microscopy. https://doi.org/10.1063/1.4864127
Cite the original work for its findings. Save a collection to share your selection of sources.