arXiv · 1710.06685
Nanometer Resolution Elemental Mapping in Graphene-based TEM Liquid Cells
Abstract
We demonstrate a new design of graphene liquid cell consisting of a thin lithographically patterned hexagonal boron nitride crystal encapsulated from both sides with graphene windows. The ultra-thin window liquid cells produced have precisely controlled volumes and thicknesses, and are robust to repeated vacuum cycling. This technology enables exciting new opportunities for liquid cell studies, providing a reliable platform for high resolution transmission electron microscope imaging and spectral mapping. The presence of water was confirmed using electron energy loss spectroscopy (EELS) via the detection of the oxygen K-edge and measuring the thickness of full and empty cells. We demonstrate the imaging capabilities of these liquid cells by tracking the dynamic motion and interactions of small metal nanoparticles with diameters of 0.5-5 nm. We further present an order of magnitude improvement in the analytical capabilities compared to previous liquid cell data, with 1 nm spatial resolution elemental mapping achievable for liquid encapsulated bimetallic nanoparticles using energy dispersive X-ray spectroscopy (EDXS).
Explore related subjects
Keep this discovery
Daniel J. Kelly, Mingwei Zhou, Nick Clark, Matthew J. Hamer, Edward A. Lewis, Alexander M. Rakowski, Sarah J. Haigh, Roman V. Gorbachev. 2017-10-18. Nanometer Resolution Elemental Mapping in Graphene-based TEM Liquid Cells. https://doi.org/10.1021/acs.nanolett.7b04713
Cite the original work for its findings. Save a collection to share your selection of sources.