arXiv · 1801.10120
Development of a new UHV/XHV pressure standard (cold atom vacuum standard)
Abstract
The National Institute of Standards and Technology has recently begun a program to develop a primary pressure standard that is based on ultra-cold atoms, covering a pressure range of 1 x 10-6 to 1 x 10-10 Pa and possibly lower. These pressures correspond to the entire ultra-high vacuum range and extend into the extreme-high vacuum. This cold-atom vacuum standard (CAVS) is both a primary standard and absolute sensor of vacuum. The CAVS is based on the loss of cold, sensor atoms (such as the alkali-metal lithium) from a magnetic trap due to collisions with the background gas (primarily H2) in the vacuum. The pressure is determined from a thermally-averaged collision cross section, which is a fundamental atomic property, and the measured loss rate. The CAVS is primary because it will use collision cross sections determined from ab initio calculations for the Li + H2 system. Primary traceability is transferred to other systems of interest using sensitivity coefficients.
Explore related subjects
Keep this discovery
Julia Scherschligt, James A Fedchak, Daniel S Barker, Stephen Eckel, Nikolai Klimov, Constantinos Makrides, Eite Tiesinga. 2018-01-30. Development of a new UHV/XHV pressure standard (cold atom vacuum standard). https://doi.org/10.1088/1681-7575%2Faa8a7b