arXiv · 1804.10838
Numerical Simulations to Evaluate and Compare the Performances of Existing and Novel Degrader Materials for Proton Therapy
Abstract
The performance of the energy degrader in terms of beam properties directly impacts the design and cost of cyclotron-based proton therapy centers. The aim of this study is to evaluate the performances of different existing and novel degrader materials. The quantitative estimate is based on detailed GEANT4 simulations that analyze the beam-matter interaction and provide a determination of the beam emittance increase and transmission. Comparisons between existing (aluminum, graphite, beryllium) and novel (boron carbide and diamond) degrader materials are provided and evaluated against semi-analytical models of multiple Coulomb scattering. The results showing a potential in emittance reduction for novel materials are presented and discussed in detail.
Explore related subjects
Keep this discovery
Robin Tesse, Alain Dubus, Nicolas Pauly, Cédric Hernalsteens, Willem Kleeven, Frederic Stichelbaut. 2018-04-28. Numerical Simulations to Evaluate and Compare the Performances of Existing and Novel Degrader Materials for Proton Therapy. https://doi.org/10.1088/1742-6596%2F1067%2F9%2F092001
Cite the original work for its findings. Save a collection to share your selection of sources.