arXiv · 1904.10779
Resolution limits of extrinsic Fabry-Perot interferometric displacement sensors utilizing wavelength scanning interrogation
Abstract
The factors limiting the resolution of displacement sensor based on extrinsic Fabry-Perot interferometer were studied. An analytical model giving the dependency of EFPI resolution on the parameters of an optical setup and a sensor interrogator was developed. The proposed model enables one to either estimate the limit of possible resolution achievable with a given setup, or to derive the requirements for optical elements and/or a sensor interrogator necessary for attaining the desired sensor resolution. An experiment supporting the analytical derivations was performed, demonstrating a large dynamic measurement range (with cavity length from tens microns to 5 mm), a high baseline resolution (from 14 pm) and a good agreement with the model.
Explore related subjects
Keep this discovery
Nikolai Ushakov, Leonid Liokumovich. 2019-04-24. Resolution limits of extrinsic Fabry-Perot interferometric displacement sensors utilizing wavelength scanning interrogation. https://doi.org/10.1364/ao.53.005092
Cite the original work for its findings. Save a collection to share your selection of sources.