Subwavelength beam focusing using a ball-shaped lens for ultrasound imaging
Abstract
Ultrasonic superresolution images can be generated by means of (super) focusing acoustic beams to subwavelength dimensions or using algorithm-based methods. Here, we demonstrate that ultrasonic pulses which are superfocused by a ball-shaped lens can be used to produce superresolution images. The imaging system is comprised of a circular flat transducer with an operation frequency of $1$ MHz, and a ball lens centered in the beam axis of symmetry. The corresponding wavelength in water is ${\lambda}_0 = 1.53$ mm. The system resolution is $0.6{\lambda}_0$ in the focal plane at one wavelength away from the lens. The superresolution method is compared with a conventional ultrasonic system based on a spherically focused transducer. Our method presents twice more resolution with a shorter depth-of-field of $2{\lambda}_0$. Possible applications that take advantage of these features are discussed, as well as some limitations of the proposed technique.
Explore related subjects
Keep this discovery
Victor H. S. Santos, Everton B. Lima, Andre L. Baggio, Jose H. Lopes, Glauber T. Silva. 2020-01-13. Subwavelength beam focusing using a ball-shaped lens for ultrasound imaging. https://arxiv.org/abs/2001.04315