Accuracy of Real-Time Echo-Planar Imaging Phase Contrast MRI
Abstract
Compared with CINE phase contrast MRI (CINE-PC), echo-planar imaging phase contrast (EPI-PC) can achieve realtime quantification of blood flow, with lower SNR. In this study, the pulsating real model of the simulated cerebral vasculature was used to verify the accuracy of EPI-PC. The imaging time of EPI-PC was 62ms/image at 100*60 spatial resolution. The reconstructed EPI-PC flow curve was extracted by homemade post-processing software. After comparison with the CINE-PC flow curve, it was concluded that EPI-PC can provide an average flow with less than 3% error, and its flow curve will be similar to the CINE-PC flow curve in shape.
Explore related subjects
Keep this discovery
Pan Liu, Sidy Fall, Olivier Baledent. 2022-07-28. Accuracy of Real-Time Echo-Planar Imaging Phase Contrast MRI. https://arxiv.org/abs/2207.13950