arXiv · 2301.06250
High sensitivity silicon carbide divacancy-based thermometer
Abstract
Color centers in silicon carbide have become potentially versatile quantum sensors. Particularly, wide temperature range temperature sensing has been realized in recent years. However, the sensitivity is limited due to the short dephasing time of the color centers. In this work, we realize a high sensitivity silicon carbide divacancy-based thermometer using the thermal Carr-Purcell-Meiboom-Gill (TCPMG) method. First, the zero field splitting D of PL6 divacancy as a function of temperature is measured with a linear slope of -99.7 kHz/K. The coherence times of TCPMG pulses linearly increase with the pulse number and the longest coherence time is about 21 us, which is ten times larger than dephasing time. The corresponding temperature sensing sensitivity is 13.4 mK/Hz1/2, which is about 15 times higher than previous results. Finally, we monitor the laboratory temperature variations for 24 hours using the TCMPG pulse. The experiments pave the way for the applications of silicon carbide-based high sensitivity thermometer in the semiconductor industry, biology, and materials sciences.
Explore related subjects
Keep this discovery
Qin-Yue Luo, Shuang Zhao, Qi-Cheng Hu, Wei-Ke Quan, Zi-Qi Zhu, Jia-Jun Li, Jun-Feng Wang. 2023-01-16. High sensitivity silicon carbide divacancy-based thermometer. https://arxiv.org/abs/2301.06250
Cite the original work for its findings. Save a collection to share your selection of sources.