arXiv · 2301.07667
Alignment-based optically pumped magnetometer using a buffer gas cell
Abstract
Alignment-based optically pumped magnetometers (OPMs) are capable of measuring oscillating magnetic fields with high sensitivity in the fT/sqrt(Hz) range. Until now, alignment-based magnetometers have only used paraffin-coated vapour cells to extend the spin relaxation lifetimes of the alkali vapour. The drawback of these cells is that they are hand-blown and are therefore time-intensive, and somewhat unreliable, to produce. Buffer gas cells, on the other hand, can be manufactured on a mass scale using microfabrication techniques. We present the first demonstration of an alignment-based magnetometer using a buffer gas vapour cell containing caesium (Cs) alkali vapour and nitrogen (N2) buffer gas. The OPM is operated at 55 degrees C and we achieve a 325 fT/sqrt(Hz) sensitivity to 10 kHz oscillating magnetic fields with an 800 Hz bandwidth. The alignment-based magnetometer uses a single laser beam for optical pumping and probing and could potentially allow for more rapid commercialisation of radio-frequency OPMs, due to the robustness of the one-beam geometry and the potential for mass-scale microfabrication of buffer gas cells.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
L. M. Rushton, L. Elson, A. Meraki, K. Jensen. 2023-01-18. Alignment-based optically pumped magnetometer using a buffer gas cell. https://doi.org/10.1103/physrevapplied.19.064047
Cite the original work for its findings. Save a collection to share your selection of sources.