arXiv · 2606.13028
Comparing Commercial Depth Sensor Accuracy for Medical Applications
Abstract
Depth estimation has numerous medical and surgical applications. We benchmark four depth sensors on a porcine bone specimen, a porcine belly specimen, and a silicone kidney phantom using stylus-sampled references. These objects contain several real-world challenges, including homogeneous surfaces, specular surfaces, and subsurface scattering. The comparison includes stereo, structured-light, and time-of-flight sensors at a distance of approximately 50 cm. Specifically, the Intel RealSense D405 (Intel RealSense, United States), PMD Flexx2 (pmdtechnologies, Germany), Stereolabs ZED 2i (Stereolabs, France), and Zivid 2M+ 60 (Zivid, Norway) are compared. The Zivid 2M+ 60 performed best across all objects and metrics considered in this work. The ZED ranked second for real tissue, but last on the phantom.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Pit Henrich, Maximilian Weiherer, Franziska Hansen, Bernhard Egger, Franziska Mathis-Ullrich. 2026-06-11. Comparing Commercial Depth Sensor Accuracy for Medical Applications. https://arxiv.org/abs/2606.13028
Cite the original work for its findings. Save a collection to share your selection of sources.