arXiv · cond-mat/0503110
Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd$_2$O$_3$ as a standard
Abstract
The long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer was evaluated using time-dependent magnetization measurements of Gd$_2$O$_3$. In contrast to earlier claims, the amplitude of the drift was found not to exceed 1-1.5 K. 30 minutes after system stabilization the temperature deviation did not exceed 0.2 K and the temperature was fully stabilized in less than 3 hours.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
S. L. Bud'ko, P. C. Canfield. 2005-03-04. Evaluation of a long-time temperature drift in a commercial Quantum Design MPMS SQUID magnetometer using Gd$_2$O$_3$ as a standard. https://doi.org/10.1016/j.jmmm.2005.04.014