arXiv · physics/0412112

Effect of noise and modeling errors on the reliability of fully 3D Monte Carlo reconstruction in SPECT

Abstract

We recently demonstrated the value of reconstructing SPECT data with fully 3D Monte Carlo reconstruction (F3DMC), in terms of spatial resolution and quantification. This was shown on a small cubic phantom (64 projections 10 x 10) in some idealistic configurations. The goals of the present study were to assess the effect of noise and modeling errors on the reliability of F3DMC, to propose and evaluate strategies for reducing the noise in the projector, and to demonstrate the feasibility of F3DMC for a dataset with realistic dimensions. A small cubic phantom and a realistic Jaszczak phantom dataset were considered. Projections and projectors for both phantoms were calculated using the Monte Carlo simulation code GATE. Projectors with different statistics were considered and two methods for reducing noise in the projector were investigated: one based on principal component analysis (PCA) and the other consisting in setting small probability values to zero. Energy and spatial shifts in projection sampling with respect to projector sampling were also introduced to test F3DMC in realistic conditions. Experiments with the cubic phantom showed the importance of using simulations with high statistics for calculating the projector, and the value of filtering the projector using a PCA approach. F3DMC was shown to be robust with respect to energy shift and small spatial sampling off-set between the projector and the projections. Images of the Jaszczak phantom were successfully reconstructed and also showed promising results in terms of spatial resolution recovery and quantitative accuracy in small structures. It is concluded that the promising results of F3DMC hold on realistic data sets

Explore related subjects

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

D. Lazaro, Z. El Bitar, V. Breton, I. Buvat. 2004-12-17. Effect of noise and modeling errors on the reliability of fully 3D Monte Carlo reconstruction in SPECT. https://doi.org/10.1109/nssmic.2004.1462770

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Light-sheet microscopy in cancer pathology: current applications, clinical challenges and future perspectives

Light-sheet fluorescence microscopy (LSFM) has emerged as a powerful research technology for three-dimensional (3D) visualization of biological tissues and is increasingly applied to human cancer specimens, including optically cleared biopsies and surgical resection samples. In this Perspective article, we review current applications of LSFM in oncology, and discuss how recent developments in microscopy, tissue preparation, labelling, and computational analysis are expanding the possibilities of 3D cancer pathology. Although LSFM has demonstrated the feasibility of large-volume imaging in selected human specimens, its application remains predominantly within research and translational settings and is not currently part of routine clinical pathology. We therefore distinguish demonstrated applications from potential future uses. We discuss how LSFM could eventually contribute to selected diagnostic and surgical workflows by providing 3D information on tumour architecture, spatial heterogeneity, metastatic dissemination, and surgical margins. These applications remain prospective and will require substantial technical, clinical, regulatory, computational, and workflow validation before clinical implementation can be considered. Ultimately, LSFM may complement rather than replace conventional histopathology by extending pathological assessment from two-dimensional tissue sections towards 3D spatial analysis.

physics.med-ph↗

Quantitative Ultrasound Perfusion Biomarker: Cross-Modality Validation with Laser Speckle Imaging

Rationale and Objectives: Tooth extraction can cause alveolar bone resorption and variable healing despite socket preservation. Current perfusion assessment relies largely on visual evaluation, limiting objective monitoring of socket healing. We developed and validated a quantitative color flow ultrasound framework to reduce sensitivity to imaging settings and motion artifacts. Materials and Methods: Color flow ultrasound data were acquired longitudinally in 30 patients undergoing tooth extraction and grafting. Velocity- and power-weighted color pixel density (CPDvel and CPDpwr) were derived from complex-valued color flow (TRI) data at 0-1.5 mm beneath the epithelial surface. Gain robustness was assessed using a Quantitative Imaging Biomarker Alliance (QIBA) flow phantom. Concurrent laser speckle contrast imaging (LASCA) provided in vivo comparison. Results: Phantom velocity estimates remained stable across color gain settings of 20-100 a.u. at flow rates of 2, 4, 6, and 8 mL/s. Perfusion metrics decreased over approximately 2-30 weeks after extraction. Nonlinear fits yielded R2 values of 0.74 for CPDvel, 0.45 for CPDpwr, and 0.66 for LASCA. Exponential half-decay times were 60.3, 92.4, and 48.3 days, respectively, relative to the first post-extraction measurement. LASCA correlated linearly with ultrasound-derived velocity (R2 = 0.68) and power (R2 = 0.74; both p < 0.05). Conclusions: The framework supports noninvasive, quantitative monitoring of superficial perfusion during socket healing with reduced sensitivity to color gain. Direct analysis of TRI data provides depth-specific perfusion metrics and may improve consistency across imaging sessions. Further validation is needed to determine its ability to identify compromised healing and inform clinical decisions.

physics.med-ph↗

LesionStudio: A Patient-Informed Platform for Task-Specific Optimization of Emission Tomography Reconstructions

Purpose: For task-specific PET and SPECT reconstruction optimization, we developed LesionStudio, an open-source platform incorporating patient background heterogeneity, realistic lesion morphology and texture, and lesion-background interactions beyond conventional phantoms. Methods: LesionStudio, a Python framework built on PyTomography, inserts lesions into PET/SPECT data using modality-specific forward projection and stochastic background ablation. Clinical realism was evaluated by physician review of 30 inserted lesions in eight DLBCL patients from a 1,579-lesion FDG lymphoma library. Quantification and detectability were assessed in FDG PET using spherical and patient-derived lesions varying in size, LBR, and site, reconstructed with OSEM and BSREM. Results: Inserted lesions were visually indistinguishable from true disease (AUC = 0.49). OSEM improved the patient-derived quantification limit from ~3.7 mL at iteration 4 to ~2.0 mL at iteration 10, but with increased noise and edge artifacts. Weakly regularized BSREM (beta = 5) achieved a similar ~2.0 mL limit with stability, whereas stronger regularization increased under-recovery. Detectability was contrast-dependent: no lesions were detected at LBR 2:1, whereas detectable volume ranged from 0.70 to 3.63 mL at LBR 4:1 and decreased to 0.70 mL at LBR 6:1-8:1 for both OSEM and weakly regularized BSREM. Conclusion: Reconstruction settings can shift SUV metrics and detectability enough to affect interpretation, so no single protocol suits every task. OSEM required early stopping for detection and feature stability but more iterations for mean-SUV quantification, whereas weakly regularized BSREM matched both at one setting. LesionStudio (https://github.com/qurit/LesionStudio), with our public library of realistic lesions, provides a reproducible framework for task-specific optimization across radiopharmaceuticals and scanners.

physics.med-ph↗