SearcharxivSearch

arXiv subjects

Christoph Pfeiffer

Publications and source records attributed to Christoph Pfeiffer.

2 recordsLinked to original sources

SQUID magnetometer based on Grooved Dayem nanobridges and a flux transformer

We report noise measurements performed on a SQUID magnetometer implementing Grooved Dayem nanobridge of YBCO as weak-links. The SQUID shows magnetic flux noise as low as 10 $μΦ_0$/Hz$^{0.5}$. The magnetometer is realized by coupling the SQUID to a flux transformer with a two-level coupling scheme using a flip-chip approach. This improves the effective area of the SQUID and result in a magnetic field noise of 50 fT/Hz$^{0.5}$ at T=77 K.

cond-mat.supr-con

Grooved Dayem nanobridges as building blocks of high-performance YBa$_2$Cu$_3$O$_{7-δ}$ SQUID magnetometers

We present noise measurements performed on a YBa$_2$Cu$_3$O$_{7-δ}$ nanoscale weak-link-based magnetometer consisting of a Superconducting QUantum Interference Device (SQUID) galvanically coupled to a $3.5 \times 3.5~$mm$^2$ pick-up loop, reaching white flux noise levels and magnetic noise levels as low as $6~μΦ_0 / \sqrt{\mathrm{Hz}}$ and $100$~fT/$\sqrt{\mathrm{Hz}}$ at $T=77$~K, respectively. The low noise is achieved by introducing Grooved Dayem Bridges (GDBs), a new concept of weak-link. A fabrication technique has been developed for the realization of nanoscale grooved bridges, which substitutes standard Dayem bridge weak links. The introduction of these novel key blocks reduces the parasitic inductance of the weak links and increases the differential resistance of the SQUIDs. This greatly improves the device performance, thus resulting in a reduction of the white noise.

cond-mat.supr-con