Performance of a Large-Volume Cryogenic Pure CsI Detector for CEνNS and Low-Energy Rare-Event Searches

A cryogenic detector system based on two 3.3 kg high-purity CsI crystals was developed and characterized at approximately 95 K. Each wedge-shaped crystal was coupled to dual-ended 3-inch photomultiplier tubes (PMTs) for scintillation readout. The measured light yields were $28.7 \pm 0.9$ and $29.3 \pm 1.0$ photoelectrons per keV electron-equivalent (PE/keV$_{ee}$) for the two crystals, with corresponding energy resolutions of 7.2 \% and 7.7 \% (FWHM) at 59.6 keV. The detector demonstrated excellent spatial uniformity, low intrinsic radioactivity, and stable operation over a continuous one-month period. Optical photon simulations using Geant4 reproduced the observed light collection trends, providing guidance for detector optimization. These results establish cryogenic pure CsI as a scalable technology for low-threshold rare-event searches.

physics.ins-det↗

Characterization and Optimization of a Cryogenic Pure CsI Detector with Remarkable Light Yield and Unprecedented Energy Resolution for CLOVERS Experiment

In this study, we comprehensively characterized and optimized a cryogenic pure CsI (pCsI) detector. We utilized a {$\SI{2}{cm}\times\SI{2}{cm}\times\SI{2}{cm}$} cube crystal coupled with a HAMAMATSU R11065 photomultiplier tube, achieving a remarkable light yield of \SI{35.2}{PE/\keV_{ee}} and an unprecedented energy resolution of \SI{6.9}{\%} at {\SI{59.54}{\keV}}. Additionally, we measured the scintillation decay time of pCsI, which was significantly shorter than that of CsI(Na) at room temperature. Furthermore, we investigated the impact of temperature, surface treatment, and crystal shape on light yield. Notably, the light yield peaked at approximately \SI{20}{\K} and remained stable within the range of \SI{70}--\SI{100}{\K}. The light yield of the polished crystals was approximately 1.5 times greater than that of the ground crystals, whereas the crystal shape exhibited minimal influence on the light yield. These results are crucial for the design of the \SI{10}{\kg} pCsI detector for the future CLOVERS (Coherent eLastic neutrinO(V)-nucleus scattERing at China Spallation Neutron Source (CSNS)) experiment.`

physics.ins-det↗