Improving the Performance of Cryogenic Calorimeters with Nonlinear Multivariate Noise Cancellation Algorithms

State-of-the-art physics experiments require high-resolution, low-noise, and low-threshold detectors to achieve competitive scientific results. However, experimental environments invariably introduce sources of noise, such as electrical interference or microphonics. The sources of this environmental noise can often be monitored by adding specially designed "auxiliary devices" (e.g. microphones, accelerometers, seismometers, magnetometers, and antennae). A model can then be constructed to predict the detector noise based on the auxiliary device information, which can then be subtracted from the true detector signal. Here, we present a multivariate noise cancellation algorithm which can be used in a variety of settings to improve the performance of detectors using multiple auxiliary devices. To validate this approach, we apply it to simulated data to remove noise due to electromagnetic interference and microphonic vibrations. We then employ the algorithm to a cryogenic light detector in the laboratory and show an improvement in the detector performance. Finally, we motivate the use of nonlinear terms to better model vibrational contributions to the noise in thermal detectors. We show a further improvement in the performance of a particular channel of the CUORE detector when using the nonlinear algorithm in combination with optimal filtering techniques.

physics.ins-det

An automated system to define the optimal operating settings of cryogenic calorimeters

Cryogenic macro-calorimeters instrumented with NTD thermistors have been developed for several decades. The choice of the optimal bias current is crucial for a proper operation of these detectors, both in terms of energy resolution and stability. In this paper we present a set of automatic measurements and analysis procedures for the characterization and optimization of the working configuration of the NTD thermistors. The presented procedures were developed for CUORE, an array of 988 cryogenic macro-calorimeters instrumented with NTD thermistors that has been taking data since 2017. These procedures made it possible to characterize a large number of detectors in a reliable way. They are suitable enough to be used also in other large arrays of cryogenic detectors, such as CUPID.

physics.ins-det