Length-dependent SWIR upconversion spectral response of noncritically phase-matched KTP crystals
Noncritically phase-matched (NCPM) KTP crystals support large-aperture bulk operation, avoid spatial walk-off, and relax angular-alignment requirements, making them attractive for short-wave infrared upconversion detection. To guide crystal selection for different detection requirements, we quantitatively characterize how crystal length affects the coverage and profile of their external upconversion spectral responses. A calibrated Czerny--Turner monochromator is used to measure the responses of 0.5, 1.0, and 2.0 mm crystals and compare them with theoretical quantum-efficiency spectra calculated from the phase-matching model. With increasing crystal length, the response evolves from a broad profile with weak peak separation to a more distinct double-peak structure, accompanied by reduced coverage bandwidth. A representative pump-power measurement further yields the system-level external quantum-efficiency slope. These results clarify the trade-off between spectral coverage and wavelength selectivity for different crystal lengths and can be used to choose the crystal length for SWIR upconversion detection systems.