SearcharxivSearch

arXiv subjects

Yuji Kishimoto

Publications and source records attributed to Yuji Kishimoto.

2 recordsLinked to original sources

Development of a Time-resolved Neutron Imaging Detector Based on the μPIC Micro-Pixel Chamber

We have developed a prototype time-resolved neutron imaging detector employing a micro-pattern gaseous detector known as the micro-pixel chamber (μPIC) coupled with a field-programmable-gate-array-based data acquisition system. Our detector system combines 100μm-level spatial and sub-μs time resolutions with a low gamma sensitivity of less than 10^-12 and high data rates, making it well suited for applications in neutron radiography at high-intensity, pulsed neutron sources. In the present paper, we introduce the detector system and present several test measurements performed at NOBORU (BL10), J-PARC to demonstrate the capabilities of our prototype. We also discuss future improvements to the spatial resolution and rate performance.

physics.ins-det

Spatial resolution of a μPIC-based neutron imaging detector

We present a detailed study of the spatial resolution of our time-resolved neutron imaging detector utilizing a new neutron position reconstruction method that improves both spatial resolution and event reconstruction efficiency. Our prototype detector system, employing a micro-pattern gaseous detector known as the micro-pixel chamber (μPIC) coupled with a field-programmable-gate-array-based data acquisition system, combines 100μm-level spatial and sub-μs time resolutions with excellent gamma rejection and high data rates, making it well suited for applications in neutron radiography at high-intensity, pulsed neutron sources. From data taken at the Materials and Life Science Experimental Facility within the Japan Proton Accelerator Research Complex (J-PARC), the spatial resolution was found to be approximately Gaussian with a sigma of 103.48 +/- 0.77 μm (after correcting for beam divergence). This is a significant improvement over that achievable with our previous reconstruction method (334 +/- 13 μm), and compares well with conventional neutron imaging detectors and with other high-rate detectors currently under development. Further, a detector simulation indicates that a spatial resolution of less than 60 μm may be possible with optimization of the gas characteristics and μPIC structure. We also present an example of imaging combined with neutron resonance absorption spectroscopy.

physics.ins-det