Abstract
We report the testing of a charcoal-based Kr-83m source for use in calibrating a low background two-phase liquid xenon detector. Kr-83m atoms produced through the decay of Rb-83 are introduced into a xenon detector by flowing xenon gas past the Rb-83 source. 9.4 keV and 32.1 keV transitions from decaying 83Krm nuclei are detected through liquid xenon scintillation and ionization. The characteristics of the Kr-83m source are analyzed and shown to be appropriate for a low background liquid xenon detector. Introduction of Kr-83m allows for quick, periodic calibration of low background noble liquid detectors at low energy.
Explore related subjects
Keep this discovery
L. W. Kastens, S. Bedikian, S. B. Cahn, A. Manzur, D. N. McKinsey. 2009-12-11. A 83Krm Source for Use in Low-background Liquid Xenon Time Projection Chambers. https://doi.org/10.1088/1748-0221%2F5%2F05%2Fp05006