arXiv · 1009.1972
Modeling of beam customization devices in the pencil beam splitting algorithm for heavy charged particle radiotherapy
Abstract
A broad-beam-delivery system for heavy-charged-particle radiotherapy often employs multiple collimators and a range-compensating filter, which potentially offer complex beam customization. In treatment planning, it is however difficult for a conventional pencil-beam algorithm to deal with these structures due to beam-size growth during transport. This study aims to resolve the problem with a novel computational model. The pencil beams are initially defined at the range compensating filter with angular-acceptance correction for the upstream collimators followed by the range compensation effects. They are individually transported with possible splitting near the downstream collimator edges to deal with its fine structure. The dose distribution for a carbon-ion beam was calculated and compared with existing experimental data. The penumbra sizes of various collimator edges agreed between them to a submillimeter level. This beam-customization model will complete an accurate and efficient dose-calculation algorithm for treatment planning with heavy charged particles.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Nobuyuki Kanematsu. 2010-12-06. Modeling of beam customization devices in the pencil beam splitting algorithm for heavy charged particle radiotherapy. https://doi.org/10.1088/0031-9155%2F56%2F5%2F009
Cite the original work for its findings. Save a collection to share your selection of sources.