arXiv · 1108.5143
Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection
Abstract
We present the first operation of the Avalanche Photodiode (APD) from Hamamatsu to xenon scintillation light and to direct X-rays of 22.1 keV and 5.9 keV. A large non-linear response was observed for the direct X-ray detection. At 415 V APD bias voltage it was of about 30 % for 22.1 keV and about 45 % for 5.9 keV. The quantum efficiency for 172 nm photons has been measured to be 69 +/- 15 %.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
T. Lux, E. D. C. Freitas, F. D. Amaro, O. Ballester, G. V. Jover-Manas, C. Martin, C. M. B. Monteiro, F. Sanchez, J. Rico. 2012-01-05. Characterization of the Hamamatsu S8664 Avalanche Photodiode for X-Ray and VUV-light detection. https://doi.org/10.1016/j.nima.2012.04.086
Cite the original work for its findings. Save a collection to share your selection of sources.