arXiv · 1202.6495
High count rate γ-ray spectroscopy with LaBr3:Ce scintillation detectors
Abstract
The applicability of LaBr3:Ce detectors for high count rate γ-ray spectroscopy is investigated. A 3"x3" LaBr3:Ce detector is used in a test setup with radioactive sources to study the dependence of energy resolution and photo peak efficiency on the overall count rate in the detector. Digitized traces were recorded using a 500 MHz FADC and analysed with digital signal processing methods. In addition to standard techniques a pile-up correction method is applied to the data in order to further improve the high-rate capabilities and to reduce the losses in efficiency due to signal pile-up. It is shown, that γ-ray spectroscopy can be performed with high resolution at count rates even above 1 MHz and that the performance can be enhanced in the region between 500 kHz and 10 MHz by using pile-up correction techniques.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
B. Löher, D. Savran, E. Fiori, M. Miklavec, N. Pietralla, M. Vencelj. 2012-05-10. High count rate γ-ray spectroscopy with LaBr3:Ce scintillation detectors. https://doi.org/10.1016/j.nima.2012.05.051
Cite the original work for its findings. Save a collection to share your selection of sources.