arXiv · 1311.4779
Characterization and modeling of a low background HPGe detector
Abstract
A high efficiency, low background counting setup has been made at TIFR consisting of a special HPGe detector ($\sim$ 70%) surrounded by a low activity copper+lead shield. Detailed measurements are performed with point and extended geometry sources to obtain a complete response of the detector. An effective model of the detector has been made with GEANT4 based Monte Carlo simulations which agrees with experimental data within 5%. This setup will be used for qualification and selection of radio-pure materials to be used in a cryogenic bolometer for the study of Neutrinoless Double Beta Decay in $^{124}$Sn as well as for other rare event studies. Using this setup, radio-impurities in the rock sample from India-based Neutrino Observatory (INO) site have been estimated.
Explore related subjects
Keep this discovery
N. Dokania, V. Singh, S. Mathimalar, V. Nanal, S. Pal, R. G. Pillay. 2013-11-19. Characterization and modeling of a low background HPGe detector. https://doi.org/10.1016/j.nima.2014.01.064
Cite the original work for its findings. Save a collection to share your selection of sources.