Comparison of commercial thermoluminescent readers regarding high-dose high-temperature measurements

Abstract

Three different thermoluminescent measuring systems have been compared with respect to the differences in temperature profiles, spectral sensitivities, as well as linearity of dose response characteristics. The comparison was performed using the Harshaw 3500, Riso DA-20 and RA94 TLD readers. The instruments were tested for the readouts of highly irradiated LiF:Mg,Cu,P (MCP) TL detectors, which require readout up to 550 C, in case of doses exceeding 1 kGy. It was found that the Harshaw 3500 can be used, without any additional light attenuation, for the measurements of MCP detectors exposed to doses up to about 5 Gy. For the other two readers the upper dose limit is about 5 times lower. It was also found that the Harshaw 3500 shows the best thermal stability considering the peak maximum position. For the ultra-high doses the differences in the spectral characteristics of the applied optical filters and photomultipliers, in conjunction with an evolution of the MCP TL emission spectrum with increasing dose, significantly influence the shape of TL glow curves measured with the Riso reader. The detailed characteristic of the compared TLD readers, at high-dose high-temperature range is discussed.

Explore related subjects

Keep this discovery

BibTeXRIS

P. Bilski, W. Gieszczyk, B. Obryk, K. Hodyr. 2014-02-25. Comparison of commercial thermoluminescent readers regarding high-dose high-temperature measurements. https://arxiv.org/abs/1401.6356

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det