Sources of the systematic errors in measurements of Po-214 decay half-life time variations at the Baksan deep underground experiments
Abstract
The design changes of the Baksan low-background TAU-1 and TAU-2 set-ups allowed to improve a sensitivity of Po-214 half-life (\tau) measurements up to the 2.5 x 10^{-4} are described. Different possible sources of systematic errors influencing on the $\tau$-value are studed. An annual variation of Po-214 half-life time measurements with an amplitude of A=(6.9 \pm 3) x 10^{-4} and a phase of (\phi=93 \pm 10) days was found in a sequence of the week-collected \tau-values obtained from the TAU-2 data sample with total duration of 480 days. 24 hours' variation of the \tau-value measurements with an amplitude of A=(10.0 \pm 2.6) x 10^{-4} and phase of (\phi=1 \pm 0.5) hours was found in a solar day 1 hour step \tau-value sequence formed from the same data sample. It was found that the Po-214 half-life averaged at 480 days is equal to (163.45 \pm 0.04) mks.
Explore related subjects
Keep this discovery
E. N. Alexeyev, Yu. M. Gavrilyuk, A. M. Gangapshev, V. V. Kazalov, V. V. Kuzminov, S. I. Panasenko, S. S. Ratkevich. 2014-04-23. Sources of the systematic errors in measurements of Po-214 decay half-life time variations at the Baksan deep underground experiments. https://doi.org/10.1134/s1063779615020021