arXiv · 1407.6189
Mini-review on the temporal resolution of fluorescence imaging systems
Abstract
Imaging of rapidly occurring events in biology requires high temporal resolution. In this review article, I have tried to investigate an approximate estimate to quantify temporal resolution limit of a fluorescence imaging system. It was realised that, the temporal resolution is essentially determined by the fact that, once excited, almost all ($99.9\%$) excited molecules relax to ground state. We determined the time required for $99.9\%$ of molecules to relax is about $3τ_p = 3/{(k_f +k_{nr})\log_{10}e}$, where $k_f +k_{nr}$ is the total emission rate. To arrive at this expression we assumed that, the quantum efficiency of the detector is unity, no scattering and there are no other loses. We further discuss few microscopy technique that are capable of high temporal resolution. These technique includes, multifocal multiphoton microscopy (MMM), multiple excitation spot optical microscopy (MESO) and multiple light-sheet microscopy (MLSM).
Explore related subjects
Keep this discovery
Partha Pratim Mondal. 2014-07-24. Mini-review on the temporal resolution of fluorescence imaging systems. https://arxiv.org/abs/1407.6189
Cite the original work for its findings. Save a collection to share your selection of sources.