arXiv · 1503.02788
Retrospective Estimation of the Quality of Intensity-Modulated Radiotherapy Plans for Lung Cancer
Abstract
This study estimated the planning quality of intensity-modulated radiotherapy in 42 lung cancer cases to provide preliminary data for the development of a planning quality assurance algorithm. Organs in or near the thoracic cavity (ipsilateral lung, contralateral lung, heart, liver, esophagus, spinal cord, and bronchus) were selected as organs at risk (OARs). Radiotherapy plans were compared using the conformity index (CI), coverage index (CVI), and homogeneity index (HI) of the planning target volume (PTV), OAR-PTV distance and OAR-PTV overlap volume, and the V10Gy, V20Gy, and equivalent uniform dose (EUD) of the OARs. The CI, CVI, and HI of the PTV were 0.54 - 0.89 , 0.90 - 1.00 , and 0.11 - 0.41, respectively. The mean EUDs (V10Gy, V20Gy) of the ipsilateral lung, contralateral lung, esophagus, cord, liver, heart, and bronchus were 8.07 Gy (28.06, 13.17), 2.59 Gy (6.53, 1.18), 7.02 Gy (26.17, 12.32), 3.56 Gy (13.56, 4.48), 0.72 Gy (2.15, 0.91), 5.14 Gy (19.68, 8.62), and 10.56 Gy (36.08, 19.79), respectively. EUDs tended to decrease as OAR-PTV distance increased and OAR-PTV overlap volume decreased. Because the plans in this study were from a single department, relatively few people were involved in treatment planning. Differences in treatment results for a given patient would be much more pronounced if many departments were involved.
Explore related subjects
Keep this discovery
Jihye Koo, Myonggeun Yoon, Weon Kuu Chung, Dong Wook Kim. 2015-03-10. Retrospective Estimation of the Quality of Intensity-Modulated Radiotherapy Plans for Lung Cancer. https://arxiv.org/abs/1503.02788
Cite the original work for its findings. Save a collection to share your selection of sources.