arXiv · 1507.07316
SOI pixel circuits with synchronized TMC for time-of-flight stigmatic imaging mass spectrometry
Abstract
We propose SOI pixel circuits with a synchronized time memory cell (TMC) for time-of-flight stigmatic imaging mass spectrometry. The circuits simultaneously detect the position and the fine/coarse flight time of an ion for the MALDI-ToF mass spectrometer. We discuss the circuit design and present the simulation results of a prototype detector comprised of a 32 x 32 pixel array in which each pixel pitch is 40 um and the time resolution is a minimum of 1 ns. The results of transient analysis demonstrate the fully correct synchronous operation at a 100-MHz clock frequency and simultaneous 32-word SRAM writing.
Explore related subjects
Keep this discovery
Kaori Watanabe, Masayuki Ikebe, Youichi Fujita, Yasuo Arai, Hisanao Hazama. 2015-07-27. SOI pixel circuits with synchronized TMC for time-of-flight stigmatic imaging mass spectrometry. https://arxiv.org/abs/1507.07316
Cite the original work for its findings. Save a collection to share your selection of sources.