arXiv · 1509.09021
A 128-channel picoammeter system and its application on charged particle beam current distribution measurements
Abstract
A 128-channel picoammeter system is constructed based on instrumentation amplifiers. Taking the advantages of high electric potential and narrow bandwidth in DC energetic charged beam measurements, current resolution better than 5 fA can be achieved. Two 128-channel Faraday cup arrays are built, and are employed for ion and electron beam current distribution measurements. Tests with 60 keV O3+ ions and 2 keV electrons show that it can provide exact boundaries when a positive charged particle beam current distribution is measured.
Explore related subjects
Keep this discovery
Deyang Yu, Junliang Liu, Yingli Xue, Mingwu Zhang, Xiaohong Cai, Jianjun Hu, Jinmei Dong, Xin Li. 2015-09-30. A 128-channel picoammeter system and its application on charged particle beam current distribution measurements. https://doi.org/10.1063/1.4934849
Cite the original work for its findings. Save a collection to share your selection of sources.