arXiv · 1512.00574
Ultra-Low Noise Mechanically Cooled Germanium Detector
Abstract
Low capacitance, large volume, high purity germanium (HPGe) radiation detectors have been successfully employed in low-background physics experiments. However, some physical processes may not be detectable with existing detectors whose energy thresholds are limited by electronic noise. In this paper, methods are presented which can lower the electronic noise of these detectors. Through ultra-low vibration mechanical cooling and wire bonding of a CMOS charge sensitive preamplifier to a sub-pF p-type point contact HPGe detector, we demonstrate electronic noise levels below 40 eV-FWHM.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Paul Barton, Mark Amman, Ryan Martin, Kai Vetter. 2015-12-02. Ultra-Low Noise Mechanically Cooled Germanium Detector. https://doi.org/10.1016/j.nima.2015.12.031