arXiv · 1610.02565
High-dynamic-range water window ptychography
Abstract
Ptychographic imaging with soft X-rays, especially in the water window energy range, suffers from limited detector dynamic range that directly influences the maximum spatial resolution achievable. High-dynamic-range data can be obtained by multiple exposures. By this approach we have increased the dynamic range of a ptychographic data set by a factor of 76 and obtained diffraction signal till the corners of the detector. The real space half period resolution was improved from 50 nm for the single exposure data to 18 nm for the high-dynamic-range data.
Explore related subjects
Keep this discovery
Max Rose, Dmitry Dzhigaev, Tobias Senkbeil, Andreas R. von Gundlach, Susan Stuhr, Christoph Rumancev, Ilya Besedin, Petr Skopintsev, Jens Viefhaus, Axel Rosenhahn, Ivan A. Vartanyants. 2016-10-08. High-dynamic-range water window ptychography. https://doi.org/10.1088/1742-6596/849/1/012027