Comparative study of afterpulsing behavior and models in single photon counting avalanche photo diode detectors

Abstract

Single-photon detectors, like Avalanche Photo Diodes (APDs), have a great importance in many fields like quantum key distribution, laser ranging, florescence microscopy, etc. Afterpulsing is a typical non ideal behavior of APDs, operated in Geiger mode, which adversely affects any application which measures the number or timing of detection events. Several conflicting studies have tried to model afterpulsing behavior and link it to fundamental semiconductor physics. Here we describe experiments we performed on at least 2 different detectors from 3 manufacturers, all commonly used in quantum optics setups. We report on the timing structure of the counts. By fitting different models to these data, we found statistically significant differences between the detectors used. Furthermore, we report on the presence of high order afterpulses which are not accounted for in any of the standard models of afterpulse behavior. This might have significant implication e.g. in quantum key distribution systems, as the modeling of the devices in use will have to be part of the security analysis for some systems. Significant differences between manufacturers and even between seemingly identical detectors of the same make and batch suggest that every individual detector needs to be calibrated. We describe a reliable procedure to perform such a calibration and present our findings on three different commercial products.

Explore related subjects

Keep this discovery

BibTeXRIS

Abdul W. Ziarkash, Siddarth Koduru Joshi, Mario Stipčević, Rupert Ursin. 2017-01-12. Comparative study of afterpulsing behavior and models in single photon counting avalanche photo diode detectors. https://doi.org/10.1038/s41598-018-23398-z

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det