arXiv · 1701.05341
Three-dimensional radiation dosimetry based on optically-stimulated luminescence
Abstract
A new approach to three-dimensional (3D) dosimetry based on optically-stimulated luminescence (OSL) is presented. By embedding OSL-active particles into a transparent silicone matrix (PDMS), the well-established dosimetric properties of an OSL material are exploited in a 3D-OSL dosimeter. By investigating prototype dosimeters in standard cuvettes in combination with small test samples for OSL readers, it is shown that a sufficient transparency of the 3D-OSL material can be combined with an OSL response giving an estimated >10.000 detected photons in 1 second per 1mm3 voxel of the dosimeter at a dose of 1 Gy. The dose distribution in the 3D-OSL dosimeters can be directly read out optically without the need for subsequent reconstruction by computational inversion algorithms. The dosimeters carry the advantages known from personal-dosimetry use of OSL: the dose distribution following irradiation can be stored with minimal fading for extended periods of time, and dosimeters are reusable as they can be reset, e.g. by an intense (bleaching) light field.
Explore related subjects
Keep this discovery
Michal Sadel, Ellen Marie Høye, Peter Skyt, Ludvig Paul Muren, Jørgen Breede Baltzer Petersenand, Peter Balling. 2017-01-19. Three-dimensional radiation dosimetry based on optically-stimulated luminescence. https://doi.org/10.1088/1742-6596%2F847%2F1%2F012044
Cite the original work for its findings. Save a collection to share your selection of sources.