arXiv · 1704.00980
High-precision control of static magnetic field magnitude, orientation, and gradient using optically pumped vapour cell magnetometry
Abstract
An integrated system of hardware and software allowing precise definition of arbitrarily oriented magnetic fields up to |B| = 1 {\mu}T within a five-layer mumetal shield is described. The system is calibrated with reference to magnetic resonance observed between Zeeman states of the 6S$_{1/2}$ F = 4 $^{133}$Cs ground state. Magnetic field definition over the full 4{\pi} solid angle is demonstrated, with one-sigma tolerances in magnitude, orientation and gradient of {\delta}|B| = 0.94 nT, {\delta}{\theta} = 5.9 mrad and {\delta}$\nabla$ B = 13.0 pT/mm, respectively. This field control is used to empirically map Mx magnetometer signal amplitude as a function of the static field (B0) orientation.
Explore related subjects
Keep this discovery
S. J. Ingleby, P. F. Griffin, A. S. Arnold, M. Chouliara, E. Riis. 2017-04-04. High-precision control of static magnetic field magnitude, orientation, and gradient using optically pumped vapour cell magnetometry. https://doi.org/10.1063/1.4980159
Cite the original work for its findings. Save a collection to share your selection of sources.