arXiv · 1707.05879
A Compact Cryogenic Thermal Source for Detector Array Characterization
Abstract
We describe the design, fabrication, and validation of a cryogenically-compatible quasioptical thermal source designed to be used for characterization of detector arrays. The source is constructed using a graphite-loaded epoxy mixture that is molded into a tiled pyramidal structure. The mold is fabricated using a hardened steel template produced via a wire EDM process. The absorptive mixture is bonded to a copper backplate enabling thermalization of the entire structure. The source reflectance is measured from 30-300 GHz and compared to models.
Explore related subjects
Keep this discovery
David T. Chuss, Karwan Rostem, Edward J. Wollack, Leah Berman, Felipe Colazo, Martin DeGeorge, Kyle Helson, Marco Sagliocca. 2017-07-18. A Compact Cryogenic Thermal Source for Detector Array Characterization. https://doi.org/10.1063/1.4996751
Cite the original work for its findings. Save a collection to share your selection of sources.