arXiv · 1709.07631
Energy Dependent Features of X-ray Signals in a GridPix Detector
Abstract
We report on the calibration of an argon/isobutane (97.7 %/2.3 %)-filled GridPix detector with soft X-rays (277 eV to 8 keV) using the variable energy X-ray source of the CAST Detector Lab at CERN. We study the linearity and energy resolution of the detector using both the number of hit pixels and the total measured charge as energy measures. For the latter, the energy resolution σ_E/E is better than 10 % (20 %) for energies above 2 keV (0.5 keV). Several characteristics of the recorded events are studied.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Christoph Krieger, Jochen Kaminski, Theodoros Vafeiadis, Klaus Desch. 2017-09-22. Energy Dependent Features of X-ray Signals in a GridPix Detector. https://doi.org/10.1016/j.nima.2018.02.109
Cite the original work for its findings. Save a collection to share your selection of sources.