Abstract
We present measurements of light and charge signals in a dual-phase time projection chamber at electric fields varying from 10 V/cm up to 500 V/cm and at zero field using 511 keV gamma rays from a $^{22}$Na source. We determine the drift velocity, electron lifetime, diffusion constant, and light and charge yields at 511 keV as a function of the electric field. In addition, we fit the scintillation pulse shape to an effective exponential model, showing a decay time of 43.5 ns at low field that decreases to 25 ns at high fields.
Explore related subjects
Keep this discovery
E. Hogenbirk, M. P. Decowski, K. McEwan, A. P. Colijn. 2018-07-18. Field dependence of electronic recoil signals in a dual-phase liquid xenon time projection chamber. https://doi.org/10.1088/1748-0221/13/10/p10031