Breast biomecanical modeling for compression optimization in digital breast tomosynthesis

Abstract

Mammography is a specific type of breast imaging that uses low-dose X-rays to detect cancer in early stage. During the exam, the women breast is compressed between two plates until a nearly uniform breast thickness is obtained. This technique improves image quality and reduces dose but can also be the source of discomfort and sometimes pain for the patient. Therefore, alternative techniques allowing reduced breast compression is of potential interest. The aim of this work is to develop a 3D biomechanical Finite Element (FE) breast model in order to analyze various breast compression strategies and their impact on image quality and radiation dose. Large breast deformations are simulated using this FE model with ANSYS software. A particular attention is granted to the computation of the residual stress in the model due to gravity and boundary conditions (thorax anatomy, position of the patient inside the MRI machine). Previously developed biomechanical breast models use a simplified breast anatomy by modeling adipose and fibroglandular tissues only (Rajagopal et al. in Wiley Interdiscip Rev: Syst Biol Med 2:293--304, 2010). However, breast reconstruction surgery has proven the importance of suspensory ligaments and breast fasciae on breast mechanics (Lockwood in Plast Reconstr Surg 103:1411--1420, 1999). We are therefore consider using a more realistic breast anatomy by including skin, muscles, and suspensory ligaments. The breast tissues are modeled as neo-Hookean materials. A physical correct modeling of the breast requires the knowledge of the stress-free breast configuration. Here, this undeformed shape (i.e., without any residual stress) is computed using the prediction--correction iterative scheme proposed by Eiben et al. (Ann of Biomed Eng 44:154--173, 2016). The unloading procedure uses the breast configuration in prone and supine position in order to find a unique displacement vector field induced by gravitational forces. The 3D breast geometry is reconstructed from MRI images that are segmented (Yushkevich et al. in Neuroimage 31:1116--1128, 2006) to differentiate the four main tissue types. The breast volume is discretized with a hexa-dominant FE meshing tool as a unique volume. Finally, the model is evaluated by comparing the estimated breast deformations under gravity load with the experimental ones measured in three body positions: prone, supine, and oblique supine.

Explore related subjects

Keep this discovery

BibTeXRIS

Mîra Anna, Ann-Katherine Carton, Serge Muller, Yohan Payan. 2018-11-22. Breast biomecanical modeling for compression optimization in digital breast tomosynthesis. https://doi.org/10.1007/978-3-319-59764-5_4

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph