Characterization of the Si(Li) detector for Monte Carlo calculations of beta spectra
Abstract
A precise model of a Si(Li) detector ORTEC model SLP-06165P-OPT-0.5 was created for beta spectra calculations using the Monte Carlo (MC) code MCNPX. Detector parameters were determined from X-ray radiograms obtained with a film and a Timepix detector. The MC model of the detector was validated by comparison of calculated and experimental full-energy peak efficiencies in the energy range from 5 to 136 keV using a range of point-like photon emitting radionuclide standards. A comparison of measured and calculated beta spectrum of a radionuclide Pm-147 is presented.
Explore related subjects
Keep this discovery
Pavel Novotny, Pavel Dryak, Jaroslav Solc, Petr Kovar, Zdenek Vykydal. 2019-04-02. Characterization of the Si(Li) detector for Monte Carlo calculations of beta spectra. https://doi.org/10.1088/1748-0221/13/01/p01021