A preliminary study on a multi-resolution-level inverse planning algorithm for Gamma Knife radiosurgery

Abstract

Manual forward planning for GK radiosurgery is complicated and time-consuming, particularly for cases with large or irregularly shaped targets. Inverse planning eases GK planning by solving an optimization problem. However, due to the vast search space, most inverse planning algorithms have to decouple the planning process to isocenter preselection and sector duration optimization. This sequential scheme does not necessarily lead to optimal isocenter locations and hence optimal plans. In this study, we attempt to optimize the isocenter positions, beam shapes and durations simultaneously by proposing a multi-resolution-level (MRL) strategy to handle the large-scale GK optimization problem. In our approach, several rounds of optimizations were performed with a progressively increased spatial resolution for isocenter candidate selection. The isocenters selected from last round and their neighbors on a finer resolution were used as new isocenter candidates for next round of optimization. After plan optimization, shot sequencing was performed to group the optimized sectors to deliverable shots supported by GK treatment units. We have tested our algorithm on 6 GK cases previously treated in our institution (2 meningioma cases, 3 cases with single metastasis and 1 case with 6 metastases). Compared with manual planning, achieving same coverage and similar selectivity, our algorithm improved the gradient index averagely from 3.1 to 2.9 and reduced the maximum dose of brainstem from 8.0Gy to 5.6Gy. The average beam-on time was also reduced by from 103.8 mins to 87.4 mins. Our method was also compared with the inverse planning algorithm provided in Leksell GammaPlan planning system, and outperformed it with better plan quality for all the 6 cases.This preliminary study has demonstrated the effectiveness and feasibility of our MRL inverse planning approach for GK radiosurgery.

Explore related subjects

Keep this discovery

BibTeXRIS

Zhen Tian, Xiaofeng Yang, Matt Giles, Tonghe Wang, Hao Gao, Elizabeth Butker, Tian Liu, Shannon Kahn. 2019-10-08. A preliminary study on a multi-resolution-level inverse planning algorithm for Gamma Knife radiosurgery. https://doi.org/10.1002/mp.14078

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

Phase-contrast micro-CT for intra-operative breast tumour margin assessment using a microfocus x-ray source and photon-counting detector

Objective: Intra-operative tumour margin assessment during breast-conserving surgery requires rapid, high-resolution imaging of excised tissue, allowing the surgical team to take appropriate action within a single operation. This study evaluates a custom propagation-based phase-contrast micro-computed tomography (micro-CT) system designed to meet these clinical constraints without specialised optical elements. Methods: The experimental setup pairs a microfocus x-ray source with a photon-counting detector in a cone-beam geometry. We explore how the spatial coherence of the source can provide propagation-based phase contrast -- with no additional specialised optical elements -- and balance this against maximising the x-ray flux of the cone-beam geometry. System performance was evaluated across two anode target materials and filtration configurations at various tube power settings. Imaging capabilities were validated using anthropomorphic breast tissue phantoms and a formalin-fixed paraffin-embedded (FFPE) breast tissue specimen, with reconstructions compared against gold-standard histology. Results: An unfiltered tungsten target operated at 40 kVp yielded optimal image quality. The optimised system achieved high-resolution CT reconstructions of a 5 cm diameter sample with an isotropic voxel size of 40.7 $\upmu\text{m}$ in a scan time of 12 minutes. Reconstructed volumes demonstrated strong visual correlation with corresponding histology slides. Conclusion: Combining a microfocus source with a photon-counting detector enables high-resolution, phase-contrast micro-CT within a clinically viable timeframe, demonstrating strong potential for intra-operative margin assessment.

physics.med-ph

Understanding Search and Decision Errors in Liver Metastasis Detection and the Effects of Lower Radiation Dose

The detection performance of liver metastases decreases with the reduction of radiation dose, but misses are heterogeneous. Previous eye tracking work has characterized missed metastases into two categories: search errors i.e., the eyes never land on the lesion, and decision errors i.e., the lesion is seen but not recognized as malignant. We integrated three prior reader studies to answer this question. In all studies, radiologists interpreted the same set of 40 contrast enhanced abdominal CT exams containing 91 liver metastases whose locations had been previously marked. In two studies, the workstation recorded their gaze and eye movements. Using eye dwell times, metastases were classified as search-error-dominant (majority of misses had <2 sec gaze time) or decision-error-dominant (>2 sec gaze time). In the third study, exams were interpreted both at 120 and 200 quality reference mAs (QRM) by ten radiologists. The third study did not include eye tracking. Out of 91 liver metastases, we excluded 16 that were never missed in the eye tracking studies and used 75 liver metastases for the present study.

physics.med-ph

Develop and Optimize 5DCT Imaging Simulation and Reconstruction Methods

Purpose: To develop and optimize a 5DCT (3D + cardiac phase + respiratory phase) imaging simulation and reconstruction pipeline, and to compare two sinogram-space interpolation methods for reconstructing images at arbitrary combinations of cardiac and respiratory phase. Methods: Helical CT projections were simulated from the 4D XCAT phantom across a range of cardiac and respiratory motion states, with Poisson and electronic noise added. Ground-truth-matched volumes were generated at 5 cardiac phases and 10 respiratory amplitudes (50 total phase combinations). Because acquired projections are sparsely and unevenly distributed across this joint phase space, each target slice was reconstructed by interpolating rebinned sinogram rows to the target cardiac phase and respiratory amplitude, using either 2D scattered barycentric interpolation or 2D scattered local linear interpolation with a circular kernel for cardiac phase. Reconstructed volumes were compared to phantom ground truth using mean absolute error (MAE), and to conventional respiratory-gated 4DCT (r4DCT) reconstructed from the same simulated data. Results: Both interpolation methods eliminated the severe axial misalignment artifacts present when helical projections were reconstructed without phase-space interpolation. Local linear interpolation achieved lower MAE than barycentric interpolation across most tested conditions, with the largest improvement at low pitch. The 5DCT pipeline also produced respiratory-only volumes with fewer residual cardiac-motion artifacts than conventional r4DCT reconstructed from the same projection data, including at standard clinical pitch (0.1). Conclusions: 5DCT reconstruction using sinogram-space interpolation is feasible and can jointly resolve cardiac and respiratory motion with better accuracy than conventional 4DCT reconstruction.

physics.med-ph