Characterization of High Temperature Calibration Bath through Stability and Uniformity Tests with Data Acquisition using Standard Platinum Resistance Thermometer and Precision Multimeter

Abstract

Calibration is one area in measurement where our country is still in the process of improvement. The absence of confirmed approaches and actions for identifying the features of calibration baths by numerous laboratories disallowed them to assess more the finest measurement ability and develop their accurateness. In recent advances in technology, there is now a way of providing the best measurement capability in temperature calibration by providing validated methods in characterizing calibration baths that contributes to the uncertainty of measurement of the calibrated thermometer and is therefore important to measure the extent of its contribution to the final uncertainty. This study adapted the process from obtainable and current methods. Platinum Resistance Thermometer (PRT-Isotech) and Standard Platinum Resistance Thermometer (SPRT-Hart) which are interfaced to Precision Digital Multimeters were used for measurements. Quite a few measurements were made at varying time durations implemented on several known positions in an orderly measurement form which enclosed the whole workspace. The uniformities and stabilities of the calibration bath were computed by getting the range of the Minimum Difference with the Maximum Difference taken at every established temperature point. Outcomes of measurements exhibited that the calibration bath stability at every temperature setting are steadier at lesser temperatures and have a tendency to rise as the temperature approaches 250oC. The liquid bath was further unvarying at lesser temperatures. The variances found were insignificantly distributed since the Standard Deviation ranged only from 0.005 to 0.013.

Explore related subjects

Keep this discovery

BibTeXRIS

Arlene G. Estacio, Vilma C. Pagtalunan, Rio S. Pagtalunan, Ira C. Valenzuela, Lean Karlo S. Tolentino, Jennifer C. Dela Cruz. 2019-11-13. Characterization of High Temperature Calibration Bath through Stability and Uniformity Tests with Data Acquisition using Standard Platinum Resistance Thermometer and Precision Multimeter. https://doi.org/10.30534/ijeter/2019/047102019

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det