Detection of low energy electrons by a total absorption Cherenkov spectrometer
Abstract
Presented are results of calibration, with a quasi-monochromatic electron beam, of a total absorption Cherenkov spectrometer based on the 14.8 X0 lead glass TF-1. It was found that the energy resolution of the spectrometer was 89% to 10% in the energy range of the electron beam E = 6 to 285 MeV, respectively.
Explore related subjects
Keep this discovery
V. I. Alekseev, V. A. Baskov, V. A. Dronov, A. I. L'vov, A. V. Koltsov, Yu. F. Krechetov, E. I. Malinovsky, V. V. Polyansky. 2019-11-28. Detection of low energy electrons by a total absorption Cherenkov spectrometer. https://arxiv.org/abs/1911.12608