arXiv · 2002.00543
GEANT4 simulation of a new range verification method using delayed gamma spectroscopy of a Mo-92 marker
Abstract
In this work, we propose a novel technique for in-vivo proton therapy range verification. This technique makes use of a small hadron tumour marker, $^{92}$Mo, implanted at a short known distance from the clinical treatment volume. Signals emitted from the marker during treatment can provide a direct measurement of the proton beam energy at the marker's position. Fusion-evaporation reactions between the proton beam and marker nucleus result in the emission of delayed characteristic $\gamma$ rays, which are detected off-beam for an improved signal-to-noise ratio. In order to determine the viability of this technique and to establish an experimental setup for future work, the Monte Carlo package GEANT4 was used in combination with ROOT to simulate a treatment scenario with the new method outlined in this work. These simulations show that analyzing the intensity of delayed $\gamma$ rays produced from competing reactions yields a precise measurement of the range of the proton beam relative to the marker, with sub-millimeter uncertainty.
Explore related subjects
Keep this discovery
Eva Kasanda, Christine Burbadge, Vinzenz Bildstein, Josef Turko, Artemis Spyrou, Cornelia Höhr, Dennis Mücher. 2020-02-03. GEANT4 simulation of a new range verification method using delayed gamma spectroscopy of a Mo-92 marker. https://doi.org/10.1088/1361-6560/abbd15
Cite the original work for its findings. Save a collection to share your selection of sources.