A simple, low-cost millimeter-wave (70 GHz) interferometer with a phase accuracy better than plus or minur 2 degrees, and a response time of 10 ns is described. The simplicity of this interferometer makes it ideal for measurement of electron densities in laboratory or semiconductor processing plasmas. The relative high accuracy and low cost are attained by a homodyne system with a quadrature intermediate frequency (IF) mixer, now commercially available in millimeter-wave frequency bands. The design and construction of magnetic shielding for the system isolator, which may be required around magnetically-confined plasmas, is also described.