Furnace for Inelastic X-Ray Scattering from Liquids to 1600C
Abstract
The design and implementation of a furnace for inelastic x-ray scattering from liquids with sample temperatures up to ~1600C is described. Carbon composite heaters operating in vacuum provide robust heating elements: one pair of heaters has been used for >18 days of operational time above 1500C, including 8 cycles to room temperature. High quality data has been obtained to scattering angles as low as 7 mrad in two-theta (Q<1 nm-1 at 25.7 keV) from a sample at 1560C.
Explore related subjects
Keep this discovery
Alfred Q. R. Baron, Masanori Inui, Daisuke Ishikawa, Kazuhiro Matsuda, Yukio Kajihara, Yoichi Nakajima, Kazuhiko Taguchi, Yasunori Hattori. 2020-04-30. Furnace for Inelastic X-Ray Scattering from Liquids to 1600C. https://doi.org/10.1002/pssb.202000093