Medipix3 for dosimetry and real-time beam monitoring: first tests at a 60 MeV proton therapy facility

Abstract

Charged particle therapy (CPT) is an advanced modality of radiation therapy which has grown rapidly worldwide, driven by recent developments in technology and methods of delivery. To ensure safe and high quality treatments, various instruments are used for a range of different measurements such as for quality assurance, monitoring and dosimetry purposes. With the emergence of new and enhanced delivery techniques, systems with improved capabilities are needed to exceed existing performance limitations of conventional tools. The Medipix3 is a hybrid pixel detector able to count individual protons with millisecond time resolution at clinical flux with near instant readout and count rate linearity. The system has previously demonstrated use in medical and other applications, showing wide versatility and potential for particle therapy. In this work we present measurements of the Medipix3 detector in the 60 MeV ocular proton therapy beamline at the Clatterbridge Cancer Centre, UK. The beam current and lateral beam profiles were evaluated at multiple positions in the treatment line and compared with EBT3 Gafchromic film. The recorded count rate linearity and temporal analysis of the beam structure was measured with Medipix3 across the full range of available beam intensities, up to $3.12 \times 10^{10}$ protons/s. We explore the capacity of Medipix3 to provide non-reference measurements and its applicability as a tool for dosimetry and beam monitoring for CPT. This is the first known time the performance of the Medipix3 detector technology has been tested within a clinical, high proton flux environment.

Explore related subjects

Keep this discovery

BibTeXRIS

J. S. L. Yap, N. J. S. Bal, A. Kacperek, J. Resta López, C. P. Welsch. 2021-07-06. Medipix3 for dosimetry and real-time beam monitoring: first tests at a 60 MeV proton therapy facility. https://doi.org/10.1088/1748-0221/16/11/t11001

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det