We report on the study of a detection scheme based on a Bloch separator in two-photon Bragg interferometers. We increase the spatial separation between the two output ports of the interferometer by selectively imparting 30 Bloch oscillations to one of them before their detection via time of flight. This method allows increasing the duration of the interferometer by reducing the time for discriminating the ports of the interferometer at detection. We study in detail the impact of this separator on the performance of a dual gravity sensor, and in particular on its measurement sensitivities to gravity acceleration and gravity gradients.
Explore related subjects
Keep this discovery
Raphaël Piccon, Sumit Sarkar, Sébastien Merlet, Franck Pereira dos Santos. 2022-03-07. Separating the output ports of a Bragg interferometer via velocity selective transport. https://doi.org/10.1103/physreva.106.013303