Scintillation Event Localization in Hemi-Ellipsoid Detector for SPECT, a simulation study using Geant4 Monte-Carlo

Abstract

A high sensitivity Cardiac SPECT system using curved crystals with pinhole collimation was proposed previously (Dey, IEEE TNS 2012, Bhusal et al, Med. Phys. 2019). Here, we hypothesize that a high curvature hemi-ellipsoid detector results in measurable differences in light distribution from events at different depths in the crystal. This was tested by analyzing the scintillation light in hemi-ellipsoid detector using Monte-Carlo (Geant4) and evaluation of both the localization error at detector and the back-projected errors in object-space. We used Geant4 to simulate the propagation of scintillation light in a monolithic hemi-ellipsoidal CsI crystal. A look-up table (LUT) was created to map the points inside the crystal to the expected light pattern on the crystal surface using Geant4. In thirteen zones across the crystal, gamma-rays were simulated and the resulting scintillator light intensity on the surface was captured, serving as our experimental interactions. A Poisson-statistics-based algorithm was developed to limit the search of the gamma-ray event locations into small regions of the LUT and fine-tuned by interpolating between selected LUT points by comparing the light distribution of the gamma interactions and LUT light patterns. The localized events were individually back-projected to the object mid-plane, and the errors recorded. Excluding some outliers (up to 2%), the localized errors averaged over all the zones was 0.71 (+/-0.44) mm with a worse case of 1.36 (+/-0.67) mm at the apex. When back-projected to the midplane of the object for Cardiac SPECT, the errors were <1mm and average error was 0.4(+/-0.22) mm, due to the high system magnification afforded by the detector. Thus, for our high sensitivity system, we were also able to achieve high resolution, assuming perfect pinhole collimator resolution recovery for Cardiac SPECT application.

Explore related subjects

Keep this discovery

BibTeXRIS

Hanif R. Soysal, J. Dey, W. P. Donahue, K. L. Matthews. 2022-04-22. Scintillation Event Localization in Hemi-Ellipsoid Detector for SPECT, a simulation study using Geant4 Monte-Carlo. https://doi.org/10.1088/1748-0221/17/05/p05035

Cite the original work for its findings. Save a collection to share your selection of sources.

KEEP EXPLORING

Related papers

High-Speed Semi-FE Readout Module for ATLAS MDT at HL-LHC: Design and Production-Level Characterization

The High-Luminosity upgrade of the Large Hadron Collider (HL-LHC) introduces increased demands on the ATLAS Muon Spectrometer, particularly in terms of data throughput, timing distribution and system reliability. The Phase-II Chamber Service Module (CSM) is a key component of the upgraded Monitored Drift Tube (MDT) trigger and readout system, providing a high-speed interface between the front-end electronics and the backend systems. This paper describes the design and implementation of the Phase-II CSM, together with its validation. The results show that the CSM supports two independent optical uplinks, each operating at a line rate of 10.24 Gbps, together with clock distribution and slow control in the expected operating environment. Integration with small-diameter MDT (sMDT) chambers and tests with the prototype L0MDT trigger system are also presented. The CSM boards are now in production and will be used for installation and integration during the upcoming LHC Long Shutdown.

physics.ins-det

Spectral Discrimination of Deposited Gamma-Ray Energies in a Simulated CeBr$_3$ Scintillator

We show that wavelength measurements of individual detected optical photons may provide additional information about gamma-ray energy deposited in a CeBr$_3$ crystal when the detected-photon-count distributions overlap for nearby gamma-ray energies. Monoenergetic 662 and 629 keV gammas are used in a Geant4 simulation of a $25\times25\times20~\mathrm{mm^3}$ CeBr$_3$ crystal. Assuming a light yield of $6.0\times10^4$ photons/MeV, a wavelength-independent photon-detection efficiency of 30%, and a wavelength resolution of $\sigma_{\lambda}=40$ nm, we find that the fraction of photons reconstructed above 385 nm gives an event-level separation of $\sim$ 2 standard deviations between the 662 and 629 keV event populations selected within the same $\sim$ 1%-wide detected-photon-count interval. No timing or reconstructed interaction-position information is used. The result demonstrates, within the present simulation model, that event-dependent optical spectra can retain energy information beyond an undifferentiated photon count.

physics.ins-det

Operation of a negative ion gas time projection chamber without electronegative fill gases

The high fidelity reconstruction of particle tracks in micropatterned gaseous time projection chambers renders this technology ideal for future rare-event searches, including direction-sensitive dark matter experiments. Large drift distances are typically required for such experiments, so that the overall spatial resolution is limited by diffusion. Negative ion drift exhibits lower diffusion than electron drift and is thus an attractive option for realising a large-scale detector. The use of electronegative gases to create negative ions introduces technical challenges, most notably a reduction in gain when compared to conventional gas mixtures. In this study, we demonstrate a new method for negative ion generation via dissociative electron attachment using the conventional molecular fill gas CF$_4$. Our optical measurements of negative ion drift indicate electron attachment lengths of $<$1 mm and comparable gain to electron avalanches. The individual negative ion avalanches were also time-resolved, allowing the number of ions reaching the readout to be counted. We measure an improved energy resolution by single ion counting, relative to an integrated electron avalanche signal measured under identical gain conditions.

physics.ins-det