arXiv · 2211.00413
Construction of a radiation beam scanner and investigation of volume averaging correction factor effects on beam-profile
Abstract
In radiation beam-profile measurements, an accurate positioning of the detector with high position resolution is essential. For this purpose, we built a scanning device capable of moving a detector in three dimensions using mainly parts from a commercial 3D-printer. The accuracy and repeatability of movement was tested with caliper, laser displacement sensor, and repeated $^{60}$Co beam-profile measurements in a water phantom. The results from the caliper and the sensor showed position accuracy for the scanner to be better than $\pm$150 $μ$m. The standard deviation of the error in position from laser sensor measurements was approximately 30 $μ$m and the beam profile scans showed a maximum deviation from the mean position of 50 $μ$m. The effect of volume averaging correction factors on $^{60}$Co beam-profile was investigated with two different sized ionization chambers. The differences in the profiles were reduced significantly after applying the correction factors.
Explore related subjects
Keep this discovery
Explore connections, maps & timelines
Joonas Tikkanen, Jan Morelius, Teemu Siiskonen. 2022-11-01. Construction of a radiation beam scanner and investigation of volume averaging correction factor effects on beam-profile. https://arxiv.org/abs/2211.00413